METER BARO Module

BARO INTEGRATOR GUIDE
SENSOR DESCRIPTION
Ang BARO Module ay isang tumpak na barometer upang matumbasan ang matric na potensyal na mga sukat ng TEROS 31 at TEROS 32 tensiometer. Ang BARO Module ay maaaring gamitin bilang isang standalone sensor upang mabayaran ang isa o higit pang mga tensiometer sa isang lugar ng pagsukat, o bilang isang digital/analog converter upang mabayaran ang isang konektadong halaga ng TEROS 31 o TEROS 32 at i-convert ang signal ng SDI-12 sa isang analog voltage output (8-pin na bersyon lamang). Ang kumbinasyon ng BARO Module at TEROS 32 ay maaaring gamitin bilang kapalit ng T8 tensiometer. Para sa mas detalyadong paglalarawan kung paano gumagawa ng mga sukat ang sensor na ito, sumangguni sa BARO Module User Manual.

MGA APLIKASYON
- Barometric pressure measurement
- Barometric compensation of matric potential measurements
- Digital/analog converter for directly connected TEROS 31 and TEROS 32 tensiometers
- Appropriate for non-METER data loggers to connect TEROS 31 and TEROS 32
ADVANTAGES
- Ang digital sensor ay nakikipag-ugnayan ng maraming sukat sa isang serial interface
- Mababang-input voltage mga kinakailangan
- Sinusuportahan ng low-power na disenyo ang mga data logger na pinapatakbo ng baterya
- SDI-12, Modbus RTU or tensio LINK serial communications protocol supported
- Analog output supported (only 8-pin version)
ESPISIPIKASYON
| MGA ESPISIPIKASYON SA PAGSUKAT | |
| Barometric Pressure | |
| Saklaw | + 65 kPa hanggang +105 kPa |
| Resolusyon | ± 0.0012 kPa |
| Katumpakan | ± 0.05kPa |
| Temperatura | |
| Saklaw | -30 hanggang + 60 °C |
| Resolusyon | ± 0.01 °C |
| Katumpakan | ± 0.5 °C |
| MGA ESPISIPIKASYON NG KOMUNIKASYON | |
| Output | |
| Analog Output (8-pin connector only)0 to 2,000 mV (default)0 to 1,000 mV (configurable with tensio VIEW) | |
| Digital OutputSDI-12 communications protocol tensio LINK communication protocol Modbus RTU communication protocol | |
| Pagkatugma ng Data Logger | |
| Analog Output Anumang data acquisition system na may kakayahang ilipat ang 3.6- to 28-VDC excitation at single-ended o differential voltage pagsukat sa mas malaki kaysa o katumbas ng 12-bit na resolution. | |
| Digital Output Anumang data acquisition system na may kakayahang 3.6- hanggang 28-VDC excitation at RS-485 Modbus o SDI-12 na komunikasyon. | |
| PISIKAL NA ESPISIPIKASYON | |
| Mga sukat | |
| Ang haba | 80 mm (3.15 in) |
| Lapad | 29 mm (1.14 in) |
| taas | 30 mm (1.18 in) |
| Haba ng Cable | |
| 1.5 m (standard)NOTE: Contact Customer Support if a nonstandard cable length is needed. | |
| Mga Uri ng Konektor | |
| 4-pin and 8-pin M12 plug connector or stripped and tinned wires | |
| PAGSUNOD | |
| EM ISO/IEC 17050:2010 (CE Mark) | |
KATABAS NA URI AT URI NG KONEKSIYON
Refer to Figure 2 to connect the BARO Module to a data logger. Figure 2 provides a low-impedance variant of the recommended SDI-12 specification.



GABAY SA PAGSASAMA NG MODULE NG BARO 



MGA PAG-IINGAT
Ang mga METER sensor ay binuo sa pinakamataas na pamantayan, ngunit ang maling paggamit, hindi wastong proteksyon, o hindi wastong pag-install ay maaaring makapinsala sa sensor at posibleng mawalan ng warranty. Bago isama ang mga sensor sa isang network ng sensor, sundin ang mga inirerekomendang tagubilin sa pag-install at magpatupad ng mga pag-iingat upang maprotektahan ang sensor mula sa nakakapinsalang interference.
MGA KOMUNIKASYON SA SENSOR
METER digital sensors feature a serial interface with shared receive and transmit signals for communicating sensor measurements on the data wire. The sensor supports SDI-12, tensio LINK, and Modbus over RS-485 two-wire. The sensor automatically detects the interface and protocol which is being used. Each protocol has implementation advantages at mga hamon. Mangyaring makipag-ugnayan sa Customer Support ng METER kung hindi halata ang pagpili ng protocol para sa gustong aplikasyon.
- SDI-12 PANIMULA
Ang SDI-12 ay isang protocol na nakabatay sa pamantayan para sa pag-interfacing ng mga sensor sa mga data logger at data acquisition equipment. Ang maraming sensor na may mga natatanging address ay maaaring magbahagi ng karaniwang 3-wire bus (power, ground, at data). Ang two-way na komunikasyon sa pagitan ng sensor at logger ay posible sa pamamagitan ng pagbabahagi ng linya ng data para sa pagpapadala at pagtanggap gaya ng tinukoy ng pamantayan. Ang mga sukat ng sensor ay na-trigger ng protocol command. Ang SDI-12 protocol ay nangangailangan ng isang natatanging alphanumeric sensor address para sa bawat sensor sa bus upang ang isang data logger ay makapagpadala ng mga command sa at makatanggap ng mga pagbabasa mula sa mga partikular na sensor.
I-download ang SDI-12 Specification v1.3 para matuto pa tungkol sa SDI-12 protocol. - RS-485 PANIMULA
Ang RS-485 ay isang matibay na pisikal na koneksyon ng bus upang ikonekta ang maraming device sa isang bus. Ito ay may kakayahang gumamit ng napakahabang distansya ng cable sa ilalim ng malupit na kapaligiran. Sa halip na SDI-12, gumagamit ang RS-485 ng dalawang nakalaang wire para sa signal ng data. Nagbibigay-daan ito sa paggamit ng mas mahabang mga cable at mas insensitive sa panghihimasok mula sa labas ng mga pinagmumulan, dahil ang signal ay nauugnay sa iba't ibang mga wire, at ang supply ng mga alon ay hindi nakakaimpluwensya sa signal ng data. Tingnan ang Wikipedia para sa higit pang mga detalye sa RS-485. - TENSIOLINK RS-485 PANIMULA
Ang tensioLINK ay isang mabilis, maaasahan, pagmamay-ari na serial communications protocol na nakikipag-ugnayan sa interface ng RS-485. Ginagamit ang protocol na ito upang basahin ang data at i-configure ang mga feature ng device. Nagbibigay ang METER ng tensioLINK PC USB converter at software upang direktang makipag-ugnayan sa sensor, basahin ang data, at i-update ang firmware. Mangyaring makipag-ugnayan sa Customer Support para sa higit pang impormasyon tungkol sa tensioLINK. - MODBUS RTU RS-485 PANIMULA
Ang Modbus RTU ay isang karaniwang serial communications protocol na ginagamit ng Programmable Logic Controllers (PLCs) o data loggers para makipag-ugnayan sa lahat ng uri ng digital device. Gumagana ang komunikasyon sa pisikal na koneksyon ng RS-485. Ang kumbinasyon ng RS-485 para sa pisikal na koneksyon at Modbus bilang serial communications protocol ay nagbibigay-daan sa mabilis at maaasahang paglipat ng data para sa mataas na bilang ng mga sensor na konektado sa isang serial bus wire. Gamitin ang mga sumusunod na link para sa higit pang impormasyon ng Modbus: Wikipedia at modbus.org. - INTERFACING ANG SENSOR SA ISANG COMPUTER
The serial signals and protocols supported by the sensor require some type of interface hardware to be compatible with the serial port found on most computers (or USB-to-serial adapters). There are several
SDI-12 interface adapters available in the marketplace; however, METER has not tested any of these interfaces and cannot make a recommendation as to which adapters work with METER sensors. METER data loggers and the ZSC handheld device can operate as a computer-to-sensor interface for making on-demand sensor measurements.
Ang BARO Module ay maaari ding i-configure at sukatin sa pamamagitan ng tensioLINK gamit ang METER software tensioVIEW, magagamit upang i-download sa meter.ly/software. Upang ikonekta ang isang BARO Module sa isang computer, kinakailangan ang isang tensoLINK USB converter at isang angkop na adapter cable. - METER SDI-12 IMPLEMENTATION
Kung ang isang BARO Module ay konektado sa pagitan ng TEROS 31 o 32 tensiometer, ang barometric air pressure at ang absolute pressure ng TEROS tensiometer ay mababasa sa pamamagitan ng Modbus. Ang compensated matrix potential ay mababasa rin sa pamamagitan ng Modbus.
METER sensors use a low-impedance variant of the SDI-12 standard sensor circuit (Figure 2). During the power-up time, sensors output some sensor diagnostic information and should not be communicated with until the power-up time has passed. After the power up time, the sensors are fully compatible with all commands listed in the SDI-12 Specification v1.3 except for the continuous measurement commands ( aR0 – aR9 and aRC0 – aRC9 ). M , R , and C command implementations are found on pages 8–9. Out of the factory, all METER sensors start with SDI-12 address 0 . - MGA KONSIDERASYON NG SENSOR BUS
SDI-12 sensor buses require regular checking, sensor upkeep, and sensor troubleshooting. If one sensor goes down, that may take down the whole bus even if the remaining sensors are functioning normally. Power cycling the SDI-12 bus when a sensor is failing is acceptable. METER SDI-12 sensors can be power-cycled and read on the desired measurement interval or powered continuously and commands sent when a measurement is desired based on specified communication timing. Many factors influence the effectiveness of the bus configuration. Visit metergroup.com para sa mga artikulo at virtual na seminar na naglalaman ng higit pang impormasyon.
SDI-12 CONFIGURATION
Inililista ng talahanayan 1 ang configuration ng komunikasyon ng SDI-12.
| mesa 1 SDI-12 na pagsasaayos ng komunikasyon | |
| Rate ng Baud | 1,200 |
| Simulan ang Bits | 1 |
| Mga Bit ng Data | 7 (LSB muna) |
| Mga Parity Bits | 1 (kahit) |
| Itigil ang mga Bits | 1 |
| Lohika | Baliktad (aktibong mababa) |
SDI-12 TIMING
All SDI-12 commands and responses must adhere to the format in Figure 9 on the data line. Both the command and response are preceded by an address and terminated by a carriage return and line feed combination ( <CR><LF> ) and follow the timing shown in Figure 10.


KARANIWANG UTOS ng SDI-12
Kasama sa seksyong ito ang mga talahanayan ng mga karaniwang command ng SDI-12 na kadalasang ginagamit sa isang SDI-12 system at ang mga kaukulang tugon mula sa mga METER sensor.
UTOS SA PAGKILALA ( aI! )
Ang Identification command ay maaaring gamitin upang makakuha ng iba't ibang detalyadong impormasyon tungkol sa konektadong sensor. Isang example ng utos at tugon ay ipinapakita sa Halample 1, kung saan ang utos ay naka-bold at ang tugon ay sumusunod sa utos.
Example 1 1I!113METER␣ ␣ ␣BARO␣
|
Parameter |
Nakapirming Character Ang haba | Paglalarawan |
| 1Ako! | 3 | Data logger command.Request to the sensor for information from sensor address 1 . |
| 1 | 1 | Sensor address.Prepended on all responses, this indicates which sensor on the bus is returning the following information. |
| 13 | 2 | Isinasaad na sinusuportahan ng target na sensor ang SDI-12 Specification v1.3. |
| METER ␣ ␣ ␣ | 8 | Vendor identification string.( METER and three spaces ␣ ␣ ␣ for all METER sensors) |
| BARO␣ | 6 | Sensor model string.This string is specific to the sensor type. For the BARO, the string is BARO . |
| 100 | 3 | Sensor version.This number divided by 100 is the METER sensor version (e.g., 100 is version 1.00). |
| BARO-00001 | ≤13,variable | Sensor serial number. Ito ay isang variable na haba ng field. Maaaring tanggalin ito para sa mga mas lumang sensor. |
PALITAN ANG ADDRESS COMMAND ( aAB! )
The Change Address command is used to change the sensor address to a new address. All other commands support the wildcard character as the target sensor address except for this command. All METER sensors have a default address of 0 (zero) out of the factory. Supported addresses are alphanumeric (i.e., A – Z , and 0 – 9 ). An exampAng output mula sa isang METER sensor ay ipinapakita sa Halample 2, kung saan ang utos ay naka-bold at ang tugon ay sumusunod sa utos.
Exampsa 2 1A0!0
|
Parameter |
Nakapirming Character Ang haba | Paglalarawan |
| 1A0! | 4 | Data logger command. Request to the sensor to change its address from 1 to a new address of 0 . |
| 0 | 1 | Bagong sensor address. Para sa lahat ng kasunod na command, ang bagong address na ito ay gagamitin ng target na sensor. |
PAGPAPATUPAD NG UTOS
Inililista ng mga sumusunod na talahanayan ang nauugnay na mga utos ng Pagsukat ( M ), Continuous ( R ), at Concurrent ( C ) at mga kasunod na utos ng Data ( D ), kung kinakailangan.
PAGSUKAT NA UTOS SA PAGPAPATUPAD
Measurement ( M ) commands are sent to a single sensor on the SDI-12 bus and require that subsequent Data ( D ) commands are sent to that sensor to retrieve the sensor output data before initiating communication with another sensor on the bus. Please refer to Table 2 and for an explanation of the command sequence and to Table 5 for an explanation of response parameters.
Table 2 aM! pagkakasunod-sunod ng utos
| Utos | Tugon |
| Ang command na ito ay nag-uulat ng average, naipon, o maximum na mga halaga. | |
| aM! | atttn |
| aD0! | a± ± + |
| Mga komento | When a slave TEROS tensiometer is connected, <Press> hold the barometric compensated tensiometer output. If the BARO module is used in standalone <Press> returns the current barometric pressure. |
| NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement command is processed by the sensor, a service request a <CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until seconds have passed or wait until the service request is received before sending the data commands. See the SDI-12 Specifications v1.3 | |
TANDAAN: Ang pagsukat at kaukulang mga utos ng data ay nilayon na gamitin nang pabalik-balik. Matapos maproseso ng sensor ang isang utos ng pagsukat, isang kahilingan sa serbisyo a ay ipinadala mula sa sensor na nagpapahiwatig na handa na ang pagsukat. Maghintay hanggang lumipas ang ttt segundo o maghintay hanggang matanggap ang kahilingan sa serbisyo bago ipadala ang mga utos ng data. Tingnan ang dokumento ng SDI-12 Specifications v1.3 para sa karagdagang impormasyon.
KASABAY NA PAGSUKAT UTOS SA PAGPAPATUPAD
Ang mga concurrent Measurement ( C ) na mga command ay karaniwang ginagamit sa mga sensor na nakakonekta sa isang bus. Ang mga C command para sa sensor na ito ay lumihis mula sa karaniwang C command na pagpapatupad. Una, ipadala ang C command, hintayin ang tinukoy na tagal ng oras na nakadetalye sa C command response, at pagkatapos ay gamitin ang D command para basahin ang tugon nito bago makipag-ugnayan sa isa pang sensor.
Please refer to Table 3 for an explanation of the command sequence and to Table 5 for an explanation of response parameters.
| Table 3 aC! measurement command sequence | |
| Utos | Tugon |
| Ang utos na ito ay nag-uulat ng mga agarang halaga. | |
| aC! | atttnn |
| aD0! | a± ± + |
| NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement commanc is pro- cessed by the sensor, a service request a<CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until ttt sec- onds have passed or wait until the service request is received before sending the data commands. Please see the SDI-12 Specifications v1.3 document for more information. | |
NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement commanc is pro-cessed by the sensor, a service request a<CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until ttt sec-onds have passed or wait until the service request is received before sending the data commands. Please see the SDI-12 Specifications v1.3 document for more information.
PATULOY NA PAGSUKAT NG MGA UTOS SA PAGPAPATUPAD
Continuous Measurement ( R ) commands trigger a sensor measurement and return the data automatically after the readings are completed without needing to send a D command. aR0! returns more characters in its response than the 75-character limitation called out in the SDI-12 Specification v1.3. It is recommended to use a buffer that can store at least 116 characters.
Please refer to Table 4 for an explanation of the command sequence and see Table 5 for an explanation of response parameters.
| Table 4 aR0! measurement command sequence | |
| Utos | Tugon |
| Ang command na ito ay nag-uulat ng average, naipon, o maximum na mga halaga. | |
| aR0! | a± ± + |
| TANDAAN: Ang utos na ito ay hindi sumusunod sa timing ng pagtugon ng SDI-12. Tingnan ang METER SDI-12 Implementation para sa karagdagang impormasyon. | |
NOTE: This command does not adhere to the SDI-12 response timing. See METER SDI-12 Implementation for more information.
MGA PARAMETER
Table 5 lists the parameters, unit measurement, and a description of the parameters returned in command responses for the BARO Module.
| mesa 5 Mga Paglalarawan ng Parameter | ||
| Parameter | Yunit | Paglalarawan |
| ± | — | Positibo o negatibong palatandaan na nagsasaad ng tanda ng susunod na halaga |
| a | — | SDI-12 address |
| n | — | Bilang ng mga sukat (nakapirming lapad ng 1) |
| nn | — | Bilang ng mga sukat na may leading zero kung kinakailangan (fixed width ng 2) |
| ttt | s | Aabutin ang maximum na oras ng pagsukat (nakapirming lapad na 3) |
| — | Character ng tab | |
| — | Character ng pagbabalik ng karwahe | |
| — | Line feed character | |
| — | ASCII character denoting the sensor type For BARO Module, the character is ; | |
| — | METER serial checksum | |
| — | METER 6-bit na CRC |
METER MODBUS RTU SERIAL IMPLEMENTATION
Modbus over Serial Line is specified in two versions – ASCII and RTU. BARO Modules communicate using RTU mode exclusively. The following explanation is always related to RTU. Table 6 lists the Modbus RTU communication and configuration.
| mesa 6 Modbus communication characters | |
| Rate ng Baud (bps) | 9,600 bps |
| Simulan ang Bits | 1 |
| Mga Bit ng Data | 8 (LSB muna) |
| Mga Parity Bits | 0 (wala) |
| Itigil ang mga Bits | 1 |
| Lohika | Karaniwan (aktibong mataas) |
Figure 11 shows a message in RTU format. The size of the data determines the length of the message. The format of each byte in the message has 10 bits, including Start and Stop Bit. Each byte is sent from left to right: Least Significant Bit (LSB) to Most Significant Bit (MBS). If no parity is implemented, an additional stop bit is transmitted to fill out the character frame to a full 11-bit asynchronous character.
The Modbus application layer implements a set of standard Function codes divided into three categories: Public, User-defined, and Reserved. Well-defined public function codes for BARO Modules are documented in the Modbus Organization, Inc. (modbus.org) community.
For a reliable interaction between the BARO Module and a Modbus Master, a minimum 50ms delay is required between every Modbus command sent on the RS-485 bus. An additional timeout is needed for every Modbus query; this timeout is device-specific and depends on the quantity of the polled registers. Generally, 100ms will work fine for most of the BARO Module.
SUPPORTED MODBUS FUNCTIONS
Table 7 Function Definitions
| Function Code | Aksyon | Paglalarawan |
| 01 | Basahin ang coil/port status | Binabasa ang on/off status ng discrete output(s) sa ModBusSlave |
| 02 | Basahin ang katayuan ng pag-input | Binabasa ang on/off status ng discrete input(s) sa ModBusSlave |
| 03 | Basahin ang mga hawak na rehistro | Binabasa ang binary na nilalaman ng (mga) hawak na rehistro sa ModBusSlave |
| 04 | Basahin ang mga rehistro ng input | Binabasa ang binary na nilalaman ng (mga) rehistro ng input sa ModBusSlave |
| 05 | Puwersahin ang single coil/port | Pinipilit ang isang solong coil/port sa ModBusSlave na maging on o off |
| 06 | Sumulat ng isang rehistro | Nagsusulat ng halaga sa isang holding register sa ModBusSlave |
| 15 | Force multiple coils/ports | Forces multiple coils/ports in the ModBusSlave to either on or off |
| 16 | Sumulat ng maramihang mga rehistro | Writes values into a series of holding registers in the ModBusSlave |
PAGKAKAtawan ng DATA AT MGA TALAAN NG PAGREREHISTRO
Data values (setpoint values, parameters, sensor-specific measurement values, etc.) sent to and from the BARO Module use 16-bit and 32-bit holding (or input) registers with a 4-digit address notation. The address spaces are virtually distributed in different blocks for each data type. This is an approach to the Modbus Enron implementation. Table 8 shows the four main tables used by the BARO Module with their respective access rights. Table 9 describes the sub-blocks for each different data type representation.
Pakitandaan na ang ilang Modbus dataloggers ay gumagamit ng addressing na may +1 offset. Nagdudulot ito minsan ng kalituhan at nakabatay sa isang walang bisa sa detalye ng Modbus. Kung may mga problema sa pagpapatupad ng iyong Modbus program sa datalogger palaging subukang subukan ang iba't ibang mga register offset at mga uri ng data. Ang paggamit ng kilalang halaga, tulad ng temperatura, kung saan alam kung anong halaga ang aasahan ay isang magandang kasanayan upang simulan ang pagsubok.
| Table 8 Modbus Primary Tables | |||
| Numero ng Pagrehistro | Uri ng Talahanayan | Access | Paglalarawan |
| 1xxx | Mga Discrete Output Coils | Basahin/Isulat | on/off status or setup flags for the sensor |
| 2xxx | Mga Discrete Input Contact | Basahin | mga flag ng status ng sensor |
| 3xxx | Mga Rehistro ng Analog Input | Basahin | numerical input variables mula sa sensor (aktwal na mga sukat ng sensor) |
| 4xxx | Analog Output Holding Registers | Basahin/Isulat | numerical output variable para sa sensor (parameter, setpoint value, calibration atbp.) |
Para kay example, register 3001 ay ang unang analog input register (unang data address para sa input registers). Ang numeric na value na nakaimbak dito ay isang 16-bit unsigned integer-type na variable na kumakatawan sa unang parameter ng pagsukat ng sensor (presyon ng presyon). Ang parehong parameter ng pagsukat (presyon ng halaga) ay maaaring basahin sa rehistro 3201, ngunit sa pagkakataong ito bilang isang 32-bit na floating-point na halaga na may format na Big-Endian. Kung ang Modbus Master (Datalogger o isang PLC) ay sumusuporta lamang sa 32-bit na float-values na may format na Little-Endian, maaaring basahin ng isa ang parehong parameter ng pagsukat (parehong halaga ng presyon) sa register 3301. Ang Virtual Sub-Blocks ay sinadya upang pasimplehin ang pagsisikap ng user sa pagprograma ng query ng Modbus ng mga sensor.
| mesa 9 Modbus Virtual Sub-Blocks | |||
| Numero ng Pagrehistro | Access | Sukat | Sub-Table Data Uri |
| X001-X099 | Basahin/Isulat | 16 bit | nilagdaang integer |
| X101-X199 | Basahin/Isulat | 16 bit | unsigned integer |
| X201-X299 | Basahin/Isulat | 32 bit | float Big-Endian format |
| X301-X399 | Basahin/Isulat | 32 bit | float Little-Endian format |
REGISTER MAPPING
| mesa 10 May hawak na mga Register | |
| 41000 (41001*) | Address ng Alipin ng Modbus |
| Detalyadong Paglalarawan | Read or update the sensor’s modbus address |
| Uri ng Data | Unsigned integer |
| Pinahihintulutang Saklaw | 1 – 247 |
| Yunit | – |
| Mga komento | Ang na-update na slave address ay maiimbak sa nonvolatile memory ng sensor |
| mesa 11 BARO Module Input Registers | |
| 32000 (32001*) | Soil Water Potential |
| Detalyadong Paglalarawan | Compensated tension value from tensiometer |
| Uri ng Data | 32 bit floating Big-Endian |
| Pinahihintulutang Saklaw | -200 hanggang +200 |
| Yunit | kPa |
| Mga komento | Tensiometer needs to be connected as slave |
| 32001 (32002*) | Temperatura ng Lupa |
| Detalyadong Paglalarawan | High accuracy on board temperature measurement |
| Uri ng Data | 32 bit floating Big-Endian |
| Pinahihintulutang Saklaw | -30 hanggang +60 |
| Yunit | degC |
| Mga komento | Tensiometer needs to be connected as slave |
| 32002 (32003*) | Sukat ng Sensor Voltage |
| Detalyadong Paglalarawan | On board supply voltage pagsukat |
| Uri ng Data | 32 bit floating Big-Endian |
| Pinahihintulutang Saklaw | -10 hanggang +60 |
| Yunit | Volts |
| Mga komento | – |
| 32003 (32004*) | BARO Katayuan |
| Detalyadong Paglalarawan | Binary status |
| Uri ng Data | 32 bit floating Big-Endian |
| Pinahihintulutang Saklaw | 0/1 |
| Yunit | – |
| Mga komento | – |
| 32004 (32005*) | BARO Reference Pressure |
| Detalyadong Paglalarawan | On board high accuracy barometric pressure measurement |
| Uri ng Data | 32 bit floating Big-Endian |
| Pinahihintulutang Saklaw | +70 hanggang +120 |
| Yunit | kPa |
| Mga komento | – |
| Talahanayan 11 Baro Module Input Registers (ipinagpatuloy) | |
| 32005 (32006*) | Tensiometer Presyon |
| Detalyadong Paglalarawan | Absolute pressure value from tensiometer |
| Uri ng Data | 32 bit floating Big-Endian |
| Pinahihintulutang Saklaw | -200 hanggang +200 |
| Yunit | kPa |
| Mga komento | Tensiometer needs to be connected as slave |
| 32006 (32007*) | BARO Temperatura |
| Detalyadong Paglalarawan | On board temperature measurement |
| Uri ng Data | 32 bit floating Big-Endian |
| Pinahihintulutang Saklaw | -30 hanggang +60 |
| Yunit | degC |
| Mga komento | – |
*Iniulat ng ilang device ang mga address ng rehistro ng Modbus na may offset na +1. Ito ay totoo para sa Campbell Scientific Loggers at Dataker loggers. Upang mabasa ang nais na rehistro, gamitin ang numero sa panaklong.
EXAMPLE GAMIT NG CR6 DATALOGGER AT MODBUS RTU
Ang Campbell Scientific, Inc. CR6 Measurement and Control Datalogger supports Modbus master and Modbus slave communication to integrate Modbus SCADA networks. The Modbus communications protocol facilitates the exchange of information and data between a computer/HMI software, instruments (RTUs), and Modbus-compatible sensors. The CR6 datalogger communicates exclusively in RTU mode. In a Modbus network, each slave device has a unique address. Therefore, sensor devices must be configured correctly before connecting to a Modbus Network. Addresses range from 1 to 247. Address 0 is reserved for universal broadcasts.
PAGPROGRAMME NG CR6 DATALOGGER
Ang Mga Programang tumatakbo sa CR6 (at CR1000) Logger ay nakasulat sa CRBasic, isang wikang binuo ng Campkampana Siyentipiko. Ito ay isang mataas na antas ng wika na idinisenyo upang magbigay ng isang madali ngunit lubhang nababaluktot at makapangyarihang paraan ng pagtuturo sa data logger kung paano at kailan magsasagawa ng mga sukat, magproseso ng data, at makipag-usap. Maaaring malikha ang mga programa gamit ang alinman sa ShortCut Software o i-edit gamit ang CRBasic Editor, na parehong magagamit para sa pag-download bilang mga stand-alone na application sa opisyal na Campkampana Siyentipiko weblugar (www.campbellsci.com). ShortCut Software (https://www.campbellsci.com/shortcut) CRBasic Editor (https://www.campbellsci.com/crbasiceditor)
A typical CRBasic program for a Modbus application consists of the following:
- Variables and constants declarations (public or private)
- Units declarations
- Mga parameter ng pag-configure
- Data tables declarations
- Logger Initializations
- Scan (Main Loop) with all the sensors to be quired
- Function call to the Data Tables
CR6 LOGGER RS-485 CONNECTION INTERFACE
Ang universal (U) terminal ng CR6 ay nag-aalok ng 12 Channel na kumokonekta sa halos anumang uri ng sensor. Nagbibigay ito sa CR6 ng kakayahang tumugma sa higit pang mga application at inaalis ang paggamit ng maraming panlabas na peripheral.
Ang koneksyon ng Modbus CR6 na ipinapakita sa Figure 12 ay gumagamit ng interface ng RS-485 (A/B) na naka-mount sa mga terminal (C1-C2) at (C3-C4). Ang mga interface na ito ay maaaring gumana sa Half-Duplex at Full-Duplex. Ang serial interface ng BARO Module na ginamit para sa ex na itoampAng le ay konektado sa (C1-C2) na mga terminal.
BARO Module sa CR6 Datalogger Wiring Diagram
After assigning the BARO Module a unique Modbus Slave Address, it can be wired to the CR6 logger according to Figure 12. Make sure to connect the white and black wires according to their signals, respectively, to the C1 and C2 ports—the brown wire to 12V (V+) and the blue to G (GND). To control the power supply through your program, connect the brown wire directly to one of the SW12 terminals (switched 12V outputs).
EXAMPLE PROGRAMA

SUPORTA NG CUSTOMER
HILAGANG AMERIKA
Available ang mga kinatawan ng customer service para sa mga tanong, problema, o feedback Lunes hanggang Biyernes, 7:00 am hanggang 5:00 pm Pacific time.
- Email: support.environment@metergroup.com
- sales.environment@metergroup.com
- Telepono: +1.509.332.5600
- Fax: +1.509.332.5158
- Website: metergroup.com
EUROPE
- Customer service representatives are available for questions, problems, or feedback Monday through Friday,
- 8:00 to 17:00 Central European time.
- Email: support.europe@metergroup.com
- sales.europe@metergroup.com
- Telepono: +49 89 12 66 52 0
- Fax: + 49 89 12 66 52 20
- Website: metergroup.com
Kung nakikipag-ugnayan sa METER sa pamamagitan ng email, mangyaring isama ang sumusunod na impormasyon:
- Pangalan
- Address
- Numero ng telepono
- Email address
- Serial number ng instrumento
Paglalarawan ng problema
TANDAAN: Para sa mga produktong binili sa pamamagitan ng isang distributor, mangyaring makipag-ugnayan nang direkta sa distributor para sa tulong.
KASAYSAYAN NG REBISYON
Ang sumusunod na talahanayan ay naglilista ng mga pagbabago sa dokumento.
| Rebisyon | Petsa | Mga katugmang Firmware | Paglalarawan |
| 00 | 6.2025 | 1.10 | Paunang paglabas |
FAQ
What should I do if I need a nonstandard cable length?
Contact Customer Support for assistance with nonstandard cable lengths.
How do I know which communication protocol to use for my application?
Evaluate the advantages and challenges of each protocol based on your application needs. If unsure, contact METER Customer Support for guidance.