METER-logoMETER BARO Module

METER-BARO-Module -product

BARO INTEGRATOR GUIDE

SENSOR DESCRIPTION

Ang BARO Module ay isang tumpak na barometer upang matumbasan ang matric na potensyal na mga sukat ng TEROS 31 at TEROS 32 tensiometer. Ang BARO Module ay maaaring gamitin bilang isang standalone sensor upang mabayaran ang isa o higit pang mga tensiometer sa isang lugar ng pagsukat, o bilang isang digital/analog converter upang mabayaran ang isang konektadong halaga ng TEROS 31 o TEROS 32 at i-convert ang signal ng SDI-12 sa isang analog voltage output (8-pin na bersyon lamang). Ang kumbinasyon ng BARO Module at TEROS 32 ay maaaring gamitin bilang kapalit ng T8 tensiometer. Para sa mas detalyadong paglalarawan kung paano gumagawa ng mga sukat ang sensor na ito, sumangguni sa BARO Module User Manual.

METER-BARO-Module- (1)

MGA APLIKASYON

  • Barometric pressure measurement
  • Barometric compensation of matric potential measurements
  • Digital/analog converter for directly connected TEROS 31 and TEROS 32 tensiometers
  • Appropriate for non-METER data loggers to connect TEROS 31 and TEROS 32

ADVANTAGES

  • Ang digital sensor ay nakikipag-ugnayan ng maraming sukat sa isang serial interface
  • Mababang-input voltage mga kinakailangan
  • Sinusuportahan ng low-power na disenyo ang mga data logger na pinapatakbo ng baterya
  • SDI-12, Modbus RTU or tensio LINK serial communications protocol supported
  • Analog output supported (only 8-pin version)

ESPISIPIKASYON

MGA ESPISIPIKASYON SA PAGSUKAT
Barometric Pressure
Saklaw + 65 kPa hanggang +105 kPa
Resolusyon ± 0.0012 kPa
Katumpakan ± 0.05kPa
Temperatura
Saklaw -30 hanggang + 60 °C
Resolusyon ± 0.01 °C
Katumpakan ± 0.5 °C
 MGA ESPISIPIKASYON NG KOMUNIKASYON
Output
Analog Output (8-pin connector only)0 to 2,000 mV (default)0 to 1,000 mV (configurable with tensio VIEW)
Digital OutputSDI-12 communications protocol tensio LINK communication protocol Modbus RTU communication protocol
Pagkatugma ng Data Logger
Analog Output Anumang data acquisition system na may kakayahang ilipat ang 3.6- to 28-VDC excitation at single-ended o differential voltage pagsukat sa mas malaki kaysa o katumbas ng 12-bit na resolution.
Digital Output Anumang data acquisition system na may kakayahang 3.6- hanggang 28-VDC excitation at RS-485 Modbus o SDI-12 na komunikasyon.
 PISIKAL NA ESPISIPIKASYON
Mga sukat
Ang haba 80 mm (3.15 in)
Lapad 29 mm (1.14 in)
taas 30 mm (1.18 in)
Haba ng Cable
1.5 m (standard)NOTE: Contact Customer Support if a nonstandard cable length is needed.
Mga Uri ng Konektor
4-pin and 8-pin M12 plug connector or stripped and tinned wires
 PAGSUNOD
EM ISO/IEC 17050:2010 (CE Mark)

KATABAS NA URI AT URI NG KONEKSIYON
Refer to Figure 2 to connect the BARO Module to a data logger. Figure 2 provides a low-impedance variant of the recommended SDI-12 specification.

METER-BARO-Module- (2)

METER-BARO-Module- (3)

METER-BARO-Module- (4)

GABAY SA PAGSASAMA NG MODULE NG BARO METER-BARO-Module- (5)

METER-BARO-Module- (6)

METER-BARO-Module- (7)

METER-BARO-Module- (8)

MGA PAG-IINGAT

Ang mga METER sensor ay binuo sa pinakamataas na pamantayan, ngunit ang maling paggamit, hindi wastong proteksyon, o hindi wastong pag-install ay maaaring makapinsala sa sensor at posibleng mawalan ng warranty. Bago isama ang mga sensor sa isang network ng sensor, sundin ang mga inirerekomendang tagubilin sa pag-install at magpatupad ng mga pag-iingat upang maprotektahan ang sensor mula sa nakakapinsalang interference.

MGA KOMUNIKASYON SA SENSOR
METER digital sensors feature a serial interface with shared receive and transmit signals for communicating sensor measurements on the data wire. The sensor supports SDI-12, tensio LINK, and Modbus over RS-485 two-wire. The sensor automatically detects the interface and protocol which is being used. Each protocol has implementation advantages at mga hamon. Mangyaring makipag-ugnayan sa Customer Support ng METER kung hindi halata ang pagpili ng protocol para sa gustong aplikasyon.

  • SDI-12 PANIMULA
    Ang SDI-12 ay isang protocol na nakabatay sa pamantayan para sa pag-interfacing ng mga sensor sa mga data logger at data acquisition equipment. Ang maraming sensor na may mga natatanging address ay maaaring magbahagi ng karaniwang 3-wire bus (power, ground, at data). Ang two-way na komunikasyon sa pagitan ng sensor at logger ay posible sa pamamagitan ng pagbabahagi ng linya ng data para sa pagpapadala at pagtanggap gaya ng tinukoy ng pamantayan. Ang mga sukat ng sensor ay na-trigger ng protocol command. Ang SDI-12 protocol ay nangangailangan ng isang natatanging alphanumeric sensor address para sa bawat sensor sa bus upang ang isang data logger ay makapagpadala ng mga command sa at makatanggap ng mga pagbabasa mula sa mga partikular na sensor.
    I-download ang SDI-12 Specification v1.3 para matuto pa tungkol sa SDI-12 protocol.
  • RS-485 PANIMULA
    Ang RS-485 ay isang matibay na pisikal na koneksyon ng bus upang ikonekta ang maraming device sa isang bus. Ito ay may kakayahang gumamit ng napakahabang distansya ng cable sa ilalim ng malupit na kapaligiran. Sa halip na SDI-12, gumagamit ang RS-485 ng dalawang nakalaang wire para sa signal ng data. Nagbibigay-daan ito sa paggamit ng mas mahabang mga cable at mas insensitive sa panghihimasok mula sa labas ng mga pinagmumulan, dahil ang signal ay nauugnay sa iba't ibang mga wire, at ang supply ng mga alon ay hindi nakakaimpluwensya sa signal ng data. Tingnan ang Wikipedia para sa higit pang mga detalye sa RS-485.
  • TENSIOLINK RS-485 PANIMULA
    Ang tensioLINK ay isang mabilis, maaasahan, pagmamay-ari na serial communications protocol na nakikipag-ugnayan sa interface ng RS-485. Ginagamit ang protocol na ito upang basahin ang data at i-configure ang mga feature ng device. Nagbibigay ang METER ng tensioLINK PC USB converter at software upang direktang makipag-ugnayan sa sensor, basahin ang data, at i-update ang firmware. Mangyaring makipag-ugnayan sa Customer Support para sa higit pang impormasyon tungkol sa tensioLINK.
  • MODBUS RTU RS-485 PANIMULA
    Ang Modbus RTU ay isang karaniwang serial communications protocol na ginagamit ng Programmable Logic Controllers (PLCs) o data loggers para makipag-ugnayan sa lahat ng uri ng digital device. Gumagana ang komunikasyon sa pisikal na koneksyon ng RS-485. Ang kumbinasyon ng RS-485 para sa pisikal na koneksyon at Modbus bilang serial communications protocol ay nagbibigay-daan sa mabilis at maaasahang paglipat ng data para sa mataas na bilang ng mga sensor na konektado sa isang serial bus wire. Gamitin ang mga sumusunod na link para sa higit pang impormasyon ng Modbus: Wikipedia at modbus.org.
  • INTERFACING ANG SENSOR SA ISANG COMPUTER
    The serial signals and protocols supported by the sensor require some type of interface hardware to be compatible with the serial port found on most computers (or USB-to-serial adapters). There are several
    SDI-12 interface adapters available in the marketplace; however, METER has not tested any of these interfaces and cannot make a recommendation as to which adapters work with METER sensors. METER data loggers and the ZSC handheld device can operate as a computer-to-sensor interface for making on-demand sensor measurements.
    Ang BARO Module ay maaari ding i-configure at sukatin sa pamamagitan ng tensioLINK gamit ang METER software tensioVIEW, magagamit upang i-download sa meter.ly/software. Upang ikonekta ang isang BARO Module sa isang computer, kinakailangan ang isang tensoLINK USB converter at isang angkop na adapter cable.
  • METER SDI-12 IMPLEMENTATION
    Kung ang isang BARO Module ay konektado sa pagitan ng TEROS 31 o 32 tensiometer, ang barometric air pressure at ang absolute pressure ng TEROS tensiometer ay mababasa sa pamamagitan ng Modbus. Ang compensated matrix potential ay mababasa rin sa pamamagitan ng Modbus.
    METER sensors use a low-impedance variant of the SDI-12 standard sensor circuit (Figure 2). During the power-up time, sensors output some sensor diagnostic information and should not be communicated with until the power-up time has passed. After the power up time, the sensors are fully compatible with all commands listed in the SDI-12 Specification v1.3 except for the continuous measurement commands ( aR0 – aR9 and aRC0 – aRC9 ). M , R , and C command implementations are found on pages 8–9. Out of the factory, all METER sensors start with SDI-12 address 0 .
  • MGA KONSIDERASYON NG SENSOR BUS
    SDI-12 sensor buses require regular checking, sensor upkeep, and sensor troubleshooting. If one sensor goes down, that may take down the whole bus even if the remaining sensors are functioning normally. Power cycling the SDI-12 bus when a sensor is failing is acceptable. METER SDI-12 sensors can be power-cycled and read on the desired measurement interval or powered continuously and commands sent when a measurement is desired based on specified communication timing. Many factors influence the effectiveness of the bus configuration. Visit metergroup.com para sa mga artikulo at virtual na seminar na naglalaman ng higit pang impormasyon.

SDI-12 CONFIGURATION

Inililista ng talahanayan 1 ang configuration ng komunikasyon ng SDI-12.

mesa 1      SDI-12 na pagsasaayos ng komunikasyon
Rate ng Baud 1,200
Simulan ang Bits 1
Mga Bit ng Data 7 (LSB muna)
Mga Parity Bits 1 (kahit)
Itigil ang mga Bits 1
Lohika Baliktad (aktibong mababa)

SDI-12 TIMING
All SDI-12 commands and responses must adhere to the format in Figure 9 on the data line. Both the command and response are preceded by an address and terminated by a carriage return and line feed combination ( <CR><LF> ) and follow the timing shown in Figure 10.

METER-BARO-Module- (9)

METER-BARO-Module- (10)

KARANIWANG UTOS ng SDI-12
Kasama sa seksyong ito ang mga talahanayan ng mga karaniwang command ng SDI-12 na kadalasang ginagamit sa isang SDI-12 system at ang mga kaukulang tugon mula sa mga METER sensor.

UTOS SA PAGKILALA ( aI! )
Ang Identification command ay maaaring gamitin upang makakuha ng iba't ibang detalyadong impormasyon tungkol sa konektadong sensor. Isang example ng utos at tugon ay ipinapakita sa Halample 1, kung saan ang utos ay naka-bold at ang tugon ay sumusunod sa utos.

Example 1 1I!113METER␣ ␣ ␣BARO␣

Parameter

Nakapirming Character  Ang haba  Paglalarawan
 
1Ako! 3 Data logger command.Request to the sensor for information from sensor address 1 .
1 1 Sensor address.Prepended on all responses, this indicates which sensor on the bus is returning the following information.
13 2 Isinasaad na sinusuportahan ng target na sensor ang SDI-12 Specification v1.3.
METER ␣ ␣ ␣ 8 Vendor identification string.( METER and three spaces  ␣ ␣ ␣ for all METER sensors)
BARO␣ 6 Sensor model string.This string is specific to the sensor type. For the BARO, the string is BARO .
100 3 Sensor version.This number divided by 100 is the METER sensor version (e.g., 100 is version 1.00).
BARO-00001 ≤13,variable Sensor serial number. Ito ay isang variable na haba ng field. Maaaring tanggalin ito para sa mga mas lumang sensor.

PALITAN ANG ADDRESS COMMAND ( aAB! )
The Change Address command is used to change the sensor address to a new address. All other commands support the wildcard character as the target sensor address except for this command. All METER sensors have a default address of 0 (zero) out of the factory. Supported addresses are alphanumeric (i.e., A – Z , and 0 – 9 ). An exampAng output mula sa isang METER sensor ay ipinapakita sa Halample 2, kung saan ang utos ay naka-bold at ang tugon ay sumusunod sa utos.

Exampsa 2 1A0!0

 Parameter

Nakapirming Character  Ang haba        Paglalarawan
1A0! 4 Data logger command. Request to the sensor to change its address from 1 to a new address of 0 .
0 1 Bagong sensor address. Para sa lahat ng kasunod na command, ang bagong address na ito ay gagamitin ng target na sensor.

PAGPAPATUPAD NG UTOS
Inililista ng mga sumusunod na talahanayan ang nauugnay na mga utos ng Pagsukat ( M ), Continuous ( R ), at Concurrent ( C ) at mga kasunod na utos ng Data ( D ), kung kinakailangan.

PAGSUKAT NA UTOS SA PAGPAPATUPAD
Measurement ( M ) commands are sent to a single sensor on the SDI-12 bus and require that subsequent Data ( D ) commands are sent to that sensor to retrieve the sensor output data before initiating communication with another sensor on the bus. Please refer to Table 2 and for an explanation of the command sequence and to Table 5 for an explanation of response parameters.

Table 2 aM! pagkakasunod-sunod ng utos

Utos Tugon
Ang command na ito ay nag-uulat ng average, naipon, o maximum na mga halaga.
aM! atttn
aD0! a± ± +
Mga komento When a slave TEROS tensiometer is connected, <Press> hold the barometric compensated tensiometer output. If the BARO module is used in standalone <Press> returns the current barometric pressure.
NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement command is processed by the sensor, a service request a <CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until   seconds have passed or wait until the service request is received before sending the data commands. See the SDI-12 Specifications v1.3

TANDAAN: Ang pagsukat at kaukulang mga utos ng data ay nilayon na gamitin nang pabalik-balik. Matapos maproseso ng sensor ang isang utos ng pagsukat, isang kahilingan sa serbisyo a ay ipinadala mula sa sensor na nagpapahiwatig na handa na ang pagsukat. Maghintay hanggang lumipas ang ttt segundo o maghintay hanggang matanggap ang kahilingan sa serbisyo bago ipadala ang mga utos ng data. Tingnan ang dokumento ng SDI-12 Specifications v1.3 para sa karagdagang impormasyon.

KASABAY NA PAGSUKAT UTOS SA PAGPAPATUPAD
Ang mga concurrent Measurement ( C ) na mga command ay karaniwang ginagamit sa mga sensor na nakakonekta sa isang bus. Ang mga C command para sa sensor na ito ay lumihis mula sa karaniwang C command na pagpapatupad. Una, ipadala ang C command, hintayin ang tinukoy na tagal ng oras na nakadetalye sa C command response, at pagkatapos ay gamitin ang D command para basahin ang tugon nito bago makipag-ugnayan sa isa pang sensor.

Please refer to Table 3 for an explanation of the command sequence and to Table 5 for an explanation of response parameters.

Table 3      aC! measurement command sequence
Utos Tugon
Ang utos na ito ay nag-uulat ng mga agarang halaga.
aC! atttnn
aD0! a± ± +
NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement commanc is pro- cessed by the sensor, a service request a<CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until ttt sec- onds have passed or wait until the service request is received before sending the data commands. Please see the SDI-12 Specifications v1.3 document for more information.

NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement commanc is pro-cessed by the sensor, a service request a<CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until ttt sec-onds have passed or wait until the service request is received before sending the data commands. Please see the SDI-12 Specifications v1.3 document for more information.

PATULOY NA PAGSUKAT NG MGA UTOS SA PAGPAPATUPAD
Continuous Measurement ( R ) commands trigger a sensor measurement and return the data automatically after the readings are completed without needing to send a D command. aR0! returns more characters in its response than the 75-character limitation called out in the SDI-12 Specification v1.3. It is recommended to use a buffer that can store at least 116 characters.
Please refer to Table 4 for an explanation of the command sequence and see Table 5 for an explanation of response parameters.

Table 4      aR0! measurement command sequence
Utos Tugon
Ang command na ito ay nag-uulat ng average, naipon, o maximum na mga halaga.
aR0! a± ± +
TANDAAN: Ang utos na ito ay hindi sumusunod sa timing ng pagtugon ng SDI-12. Tingnan ang METER SDI-12 Implementation para sa karagdagang impormasyon.

NOTE: This command does not adhere to the SDI-12 response timing. See METER SDI-12 Implementation for more information.

MGA PARAMETER
Table 5 lists the parameters, unit measurement, and a description of the parameters returned in command responses for the BARO Module.

mesa 5      Mga Paglalarawan ng Parameter
Parameter Yunit Paglalarawan
± Positibo o negatibong palatandaan na nagsasaad ng tanda ng susunod na halaga
a SDI-12 address
n Bilang ng mga sukat (nakapirming lapad ng 1)
nn Bilang ng mga sukat na may leading zero kung kinakailangan (fixed width ng 2)
ttt s Aabutin ang maximum na oras ng pagsukat (nakapirming lapad na 3)
Character ng tab
Character ng pagbabalik ng karwahe
Line feed character
ASCII character denoting the sensor type For BARO Module, the character is ;
METER serial checksum
METER 6-bit na CRC

METER MODBUS RTU SERIAL IMPLEMENTATION
Modbus over Serial Line is specified in two versions – ASCII and RTU. BARO Modules communicate using RTU mode exclusively. The following explanation is always related to RTU. Table 6 lists the Modbus RTU communication and configuration.

mesa 6      Modbus communication characters
Rate ng Baud (bps) 9,600 bps
Simulan ang Bits 1
Mga Bit ng Data 8 (LSB muna)
Mga Parity Bits 0 (wala)
Itigil ang mga Bits 1
Lohika Karaniwan (aktibong mataas)

METER-BARO-Module- (10)Figure 11 shows a message in RTU format. The size of the data determines the length of the message. The format of each byte in the message has 10 bits, including Start and Stop Bit. Each byte is sent from left to right: Least Significant Bit (LSB) to Most Significant Bit (MBS). If no parity is implemented, an additional stop bit is transmitted to fill out the character frame to a full 11-bit asynchronous character.

The Modbus application layer implements a set of standard Function codes divided into three categories: Public, User-defined, and Reserved. Well-defined public function codes for BARO Modules are documented in the Modbus Organization, Inc. (modbus.org) community.

For a reliable interaction between the BARO Module and a Modbus Master, a minimum 50ms delay is required between every Modbus command sent on the RS-485 bus. An additional timeout is needed for every Modbus query; this timeout is device-specific and depends on the quantity of the polled registers. Generally, 100ms will work fine for most of the BARO Module.

SUPPORTED MODBUS FUNCTIONS

Table 7 Function Definitions

Function Code Aksyon Paglalarawan
01 Basahin ang coil/port status Binabasa ang on/off status ng discrete output(s) sa ModBusSlave
02 Basahin ang katayuan ng pag-input Binabasa ang on/off status ng discrete input(s) sa ModBusSlave
03 Basahin ang mga hawak na rehistro Binabasa ang binary na nilalaman ng (mga) hawak na rehistro sa ModBusSlave
04 Basahin ang mga rehistro ng input Binabasa ang binary na nilalaman ng (mga) rehistro ng input sa ModBusSlave
05 Puwersahin ang single coil/port Pinipilit ang isang solong coil/port sa ModBusSlave na maging on o off
06 Sumulat ng isang rehistro Nagsusulat ng halaga sa isang holding register sa ModBusSlave
15 Force multiple coils/ports Forces multiple coils/ports in the ModBusSlave to either on or off
16 Sumulat ng maramihang mga rehistro Writes values into a series of holding registers in the ModBusSlave

PAGKAKAtawan ng DATA AT MGA TALAAN NG PAGREREHISTRO
Data values (setpoint values, parameters, sensor-specific measurement values, etc.) sent to and from the BARO Module use 16-bit and 32-bit holding (or input) registers with a 4-digit address notation. The address spaces are virtually distributed in different blocks for each data type. This is an approach to the Modbus Enron implementation. Table 8 shows the four main tables used by the BARO Module with their respective access rights. Table 9 describes the sub-blocks for each different data type representation.

Pakitandaan na ang ilang Modbus dataloggers ay gumagamit ng addressing na may +1 offset. Nagdudulot ito minsan ng kalituhan at nakabatay sa isang walang bisa sa detalye ng Modbus. Kung may mga problema sa pagpapatupad ng iyong Modbus program sa datalogger palaging subukang subukan ang iba't ibang mga register offset at mga uri ng data. Ang paggamit ng kilalang halaga, tulad ng temperatura, kung saan alam kung anong halaga ang aasahan ay isang magandang kasanayan upang simulan ang pagsubok.

Table 8      Modbus Primary Tables
Numero ng Pagrehistro Uri ng Talahanayan Access Paglalarawan
1xxx Mga Discrete Output Coils Basahin/Isulat on/off status or setup flags for the sensor
2xxx Mga Discrete Input Contact Basahin mga flag ng status ng sensor
3xxx Mga Rehistro ng Analog Input Basahin numerical input variables mula sa sensor (aktwal na mga sukat ng sensor)
4xxx Analog Output Holding Registers Basahin/Isulat numerical output variable para sa sensor (parameter, setpoint value, calibration atbp.)

Para kay example, register 3001 ay ang unang analog input register (unang data address para sa input registers). Ang numeric na value na nakaimbak dito ay isang 16-bit unsigned integer-type na variable na kumakatawan sa unang parameter ng pagsukat ng sensor (presyon ng presyon). Ang parehong parameter ng pagsukat (presyon ng halaga) ay maaaring basahin sa rehistro 3201, ngunit sa pagkakataong ito bilang isang 32-bit na floating-point na halaga na may format na Big-Endian. Kung ang Modbus Master (Datalogger o isang PLC) ay sumusuporta lamang sa 32-bit na float-values ​​na may format na Little-Endian, maaaring basahin ng isa ang parehong parameter ng pagsukat (parehong halaga ng presyon) sa register 3301. Ang Virtual Sub-Blocks ay sinadya upang pasimplehin ang pagsisikap ng user sa pagprograma ng query ng Modbus ng mga sensor.

mesa 9      Modbus Virtual Sub-Blocks
Numero ng Pagrehistro Access Sukat Sub-Table Data Uri
X001-X099 Basahin/Isulat 16 bit nilagdaang integer
X101-X199 Basahin/Isulat 16 bit unsigned integer
X201-X299 Basahin/Isulat 32 bit float Big-Endian format
X301-X399 Basahin/Isulat 32 bit float Little-Endian format

REGISTER MAPPING

mesa 10      May hawak na mga Register
41000 (41001*) Address ng Alipin ng Modbus
Detalyadong Paglalarawan Read or update the sensor’s modbus address
Uri ng Data Unsigned integer
Pinahihintulutang Saklaw 1 – 247
Yunit
Mga komento Ang na-update na slave address ay maiimbak sa nonvolatile memory ng sensor

mesa 11      BARO Module Input Registers
32000 (32001*) Soil Water Potential
Detalyadong Paglalarawan Compensated tension value from tensiometer
Uri ng Data 32 bit floating Big-Endian
Pinahihintulutang Saklaw -200 hanggang +200
Yunit kPa
Mga komento Tensiometer needs to be connected as slave
32001 (32002*) Temperatura ng Lupa
Detalyadong Paglalarawan High accuracy on board temperature measurement
Uri ng Data 32 bit floating Big-Endian
Pinahihintulutang Saklaw -30 hanggang +60
Yunit degC
Mga komento Tensiometer needs to be connected as slave
32002 (32003*) Sukat ng Sensor Voltage
Detalyadong Paglalarawan On board supply voltage pagsukat
Uri ng Data 32 bit floating Big-Endian
Pinahihintulutang Saklaw -10 hanggang +60
Yunit Volts
Mga komento
32003 (32004*) BARO Katayuan
Detalyadong Paglalarawan Binary status
Uri ng Data 32 bit floating Big-Endian
Pinahihintulutang Saklaw 0/1
Yunit
Mga komento
32004 (32005*) BARO Reference Pressure
Detalyadong Paglalarawan On board high accuracy barometric pressure measurement
Uri ng Data 32 bit floating Big-Endian
Pinahihintulutang Saklaw +70 hanggang +120
Yunit kPa
Mga komento
Talahanayan 11 Baro Module Input Registers (ipinagpatuloy)
32005 (32006*) Tensiometer Presyon
Detalyadong Paglalarawan Absolute pressure value from tensiometer
Uri ng Data 32 bit floating Big-Endian
Pinahihintulutang Saklaw -200 hanggang +200
Yunit kPa
Mga komento Tensiometer needs to be connected as slave
32006 (32007*) BARO Temperatura
Detalyadong Paglalarawan On board temperature measurement
Uri ng Data 32 bit floating Big-Endian
Pinahihintulutang Saklaw -30 hanggang +60
Yunit degC
Mga komento

*Iniulat ng ilang device ang mga address ng rehistro ng Modbus na may offset na +1. Ito ay totoo para sa Campbell Scientific Loggers at Dataker loggers. Upang mabasa ang nais na rehistro, gamitin ang numero sa panaklong.

EXAMPLE GAMIT NG CR6 DATALOGGER AT MODBUS RTU
Ang Campbell Scientific, Inc. CR6 Measurement and Control Datalogger supports Modbus master and Modbus slave communication to integrate Modbus SCADA networks. The Modbus communications protocol facilitates the exchange of information and data between a computer/HMI software, instruments (RTUs), and Modbus-compatible sensors. The CR6 datalogger communicates exclusively in RTU mode. In a Modbus network, each slave device has a unique address. Therefore, sensor devices must be configured correctly before connecting to a Modbus Network. Addresses range from 1 to 247. Address 0 is reserved for universal broadcasts.

PAGPROGRAMME NG CR6 DATALOGGER
Ang Mga Programang tumatakbo sa CR6 (at CR1000) Logger ay nakasulat sa CRBasic, isang wikang binuo ng Campkampana Siyentipiko. Ito ay isang mataas na antas ng wika na idinisenyo upang magbigay ng isang madali ngunit lubhang nababaluktot at makapangyarihang paraan ng pagtuturo sa data logger kung paano at kailan magsasagawa ng mga sukat, magproseso ng data, at makipag-usap. Maaaring malikha ang mga programa gamit ang alinman sa ShortCut Software o i-edit gamit ang CRBasic Editor, na parehong magagamit para sa pag-download bilang mga stand-alone na application sa opisyal na Campkampana Siyentipiko weblugar (www.campbellsci.com). ShortCut Software (https://www.campbellsci.com/shortcut) CRBasic Editor (https://www.campbellsci.com/crbasiceditor)

A typical CRBasic program for a Modbus application consists of the following:

  • Variables and constants declarations (public or private)
  • Units declarations
  • Mga parameter ng pag-configure
  • Data tables declarations
  • Logger Initializations
  • Scan (Main Loop) with all the sensors to be quired
  • Function call to the Data Tables

CR6 LOGGER RS-485 CONNECTION INTERFACE
Ang universal (U) terminal ng CR6 ay nag-aalok ng 12 Channel na kumokonekta sa halos anumang uri ng sensor. Nagbibigay ito sa CR6 ng kakayahang tumugma sa higit pang mga application at inaalis ang paggamit ng maraming panlabas na peripheral.
Ang koneksyon ng Modbus CR6 na ipinapakita sa Figure 12 ay gumagamit ng interface ng RS-485 (A/B) na naka-mount sa mga terminal (C1-C2) at (C3-C4). Ang mga interface na ito ay maaaring gumana sa Half-Duplex at Full-Duplex. Ang serial interface ng BARO Module na ginamit para sa ex na itoampAng le ay konektado sa (C1-C2) na mga terminal.

BARO Module sa CR6 Datalogger Wiring DiagramMETER-BARO-Module- (12)

After assigning the BARO Module a unique Modbus Slave Address, it can be wired to the CR6 logger according to Figure 12. Make sure to connect the white and black wires according to their signals, respectively, to the C1 and C2 ports—the brown wire to 12V (V+) and the blue to G (GND). To control the power supply through your program, connect the brown wire directly to one of the SW12 terminals (switched 12V outputs).

EXAMPLE PROGRAMA

METER-BARO-Module- (13) METER-BARO-Module- (14) METER-BARO-Module- (15)

SUPORTA NG CUSTOMER

HILAGANG AMERIKA
Available ang mga kinatawan ng customer service para sa mga tanong, problema, o feedback Lunes hanggang Biyernes, 7:00 am hanggang 5:00 pm Pacific time.

EUROPE

Kung nakikipag-ugnayan sa METER sa pamamagitan ng email, mangyaring isama ang sumusunod na impormasyon:

  • Pangalan
  • Address
  • Numero ng telepono
  • Email address
  • Serial number ng instrumento

Paglalarawan ng problema

TANDAAN: Para sa mga produktong binili sa pamamagitan ng isang distributor, mangyaring makipag-ugnayan nang direkta sa distributor para sa tulong.

KASAYSAYAN NG REBISYON
Ang sumusunod na talahanayan ay naglilista ng mga pagbabago sa dokumento.

Rebisyon Petsa Mga katugmang Firmware Paglalarawan
00 6.2025 1.10 Paunang paglabas

FAQ

What should I do if I need a nonstandard cable length?

Contact Customer Support for assistance with nonstandard cable lengths.

How do I know which communication protocol to use for my application?

Evaluate the advantages and challenges of each protocol based on your application needs. If unsure, contact METER Customer Support for guidance.